Linear Pers 2400 Health Monitor Program

Format Contact ID or 4/2
All Programming is done by Downloading, all the tech can do is add sensors
RA2400 Program 1. Connect Phone Line from Modem to PERS2400
2. Press Test Button, keypad should sound 4 beeps
3. Set RA2400 to connect then press EMERGENCY button for 5 seconds
Compatible
Wireless
DXS-21 & DXS-23 Remote DXS-71 Smoke Detector
DXS-31 Door/Window DXT-41 & DXT-42 2 Keyfobs
DSX-62 & DXS-63 Wrist Panics DXT-61A Remote Panic
DXS-54 Motion Detector
Add Sensors
(by Customer)
1. Hold CLEAR button for 5 seconds, keypad should beep 5 times
2. Activate Transmitter, keypad should sound 1 beep indicating sensor learned
(If keypad sounds 1 long beep = Sensor already learned or system full, 16 Max)
Add Sensors
(by Installer)
1. Press TEST button in back of Keypad, should hear 4 beeps
2. Press TEST button again, should hear 5 beeps
3. Activate Transmitter, keypad should sound 1 beep indicating sensor learned
(If keypad sounds 1 long beep = Sensor already learned or system full, 16 Max)
Delete Sensors 1. Press TEST button in back of Keypad, should hear 4 beeps
2. Press and HOLD TEST button again, 6 beeps = All sensors are erased