Linear Pers 2400 Health Monitor Basics

This is a Medical Alert system, sensors are used as Activity Timers
if customer doesn’t move around within a certain time (elderly)
system will sound alarm and send signal. There are NO User Codes to enter
Zones 16, all wireless
Frequency 315 Mhz
Fuse Battery Fuse 2AG 3-amp hidden in back of keypad
AC Transformer 16-Volt 28AH plugged into wall somewhere in house
Batteries
Main Panel 12-volt 1.2 amp inside keypad
Doors/Windows 2 3-volt Lithium Hearing-Aid type DL or CR 2032
Panics 2 3-volt Lithium Hearing-Aid type DL or CR 1620
Smoke Detectors 2 9-volts Alkaline
Panic Button Press and hold for 3 seconds, can be Police, Fire or Medical Panic
Arm This system cannot be Armed or Disarmed, its not a Security System
Silence Alarm Press CLEAR key to silence alarm, Trouble or clear flashing Display
Home Button Press to reset Activity Timer, hold for 5 seconds to Disable Activity Timer
(Activity Timer sends signal if sensors show no activity for amount of time)
Home Light
ON Activity Timer is ON
FLASH SLOWLY Activity Timer has 15 minutes left
FLASH QUICKLY Activity Timer has Expired
Activity Timer Press HOME button, HOME light should turn ON
(Activity Timer is pre-set by Installer for 8, 12 or 24 Hours)

ON Activity Timer is ON
FLASH SLOWLY Activity Timer has 15 minutes left
FLASH QUICKLY Activity Timer has Expired
Reset Activity
Timer
Press HOME button again or activate a sensor
Test Sensors
1. Pull Keypad up to remove from wall
2. Press TEST button on back of keypad (do not hold), keypad beeps 4 times
3. Activate each sensor, keypad should beep indicating sensor condition

1 Beep = Transmitter is OK
2 Beeps = Transmitter had an Alarm
3 Beeps = Transmitter not checking in to panel
4 Beeps = Transmitter has Low Battery
(Wireless motions require no movement for 4 minutes prior to testing)